ResMed · Auto BiPAP

ResMed AirCurve 11 VAuto

The AirCurve 11 VAuto is ResMed's current auto-adjusting bilevel machine, announced for the US on February 12, 2024. It gives a higher pressure when you breathe in and a lower one when you breathe out, with the VAuto algorithm adjusting both within limits your doctor sets (up to 25 cm H2O). It uses the AirSense 11 body, touch screen, HumidAir 11 humidifier and myAir app, and the same user guide also covers the fixed bilevel AirCurve 11 S and the AirCurve 11 ASV.

Specifications

Noise27 dBA sound pressure level with standard tubing and HumidAir 11 tub, 26 dBA with side cover, uncertainty 2 dBA
PowerAir11 90W power supply, 100 to 240 V, 50 to 60 Hz (115 V 400 Hz for aircraft), 24 V DC output; optional Air11 DC/DC converter (39231)
Weight43 oz (1229 g) with HumidAir 11 tub; 44 oz (1236 g) with side cover
DisplayColor touch screen with light sensor; MY OPTIONS, MY SLEEP VIEW and MORE menus
Data cardSD card slot behind a removable cover; SD cards available as Air 11 SD card packs
Dimensions3.72 x 10.21 x 5.45 in (94.5 x 259.4 x 138.5 mm) H x W x D with HumidAir 11 tub; 3.72 x 9.32 x 5.45 in with side cover
HumidifierIntegrated HumidAir 11 heated humidifier, 380 mL tub, Humidity Level Off or 1 to 8, Climate Control Auto or Manual with ClimateLineAir 11
ConnectivityBuilt-in cellular and Bluetooth (myAir), Airplane Mode, over-the-air software updates
Pressure rangeVAuto mode: IPAP 4 to 25 cm H2O, EPAP 4 to 25 cm H2O, pressure support 0 to 10 cm H2O, maximum 25 cm H2O; S mode: IPAP 4 to 25, EPAP 3 to 25; CPAP mode: 4 to 20 cm H2O

Key features

  • VAuto mode: automatic bilevel pressure based on ResMed's AutoSet algorithm, with fixed pressure support set by your clinician
  • Also runs CPAP and S (fixed bilevel) modes; the AirCurve 11 S is the fixed bilevel model in the same guide
  • Ramp Time Off or 5 to 45 minutes in 5-minute steps
  • Climate Control Auto or Manual with ClimateLineAir 11 heated tubing
  • SmartStart and SmartStop when enabled by your care provider
  • Mask and Tube type settings in MY OPTIONS
  • Run Warmup, Mask Fit and Device Diagnostic in the MORE menu
  • MY SLEEP VIEW shows usage, events per hour and mask seal
  • myAir app with Personal Therapy Assistant and Care Check-In for bilevel users
  • Meets FAA requirements for use on aircraft

Strengths and trade-offs

Strengths

  • Lower pressure on the out-breath can help if you struggle to exhale against CPAP, which is the reason ResMed gives for bilevel therapy
  • About 5.45 in deep and 3.72 in tall, the same compact body as the AirSense 11
  • Works with myAir, unlike AirCurve 10 Card-to-Cloud units
  • Pressure up to 25 cm H2O in bilevel modes, higher than the 20 cm H2O ceiling of AirSense machines
  • Cleanable tub option accepts drinking-quality water; the tub and outlet connector are dishwasher safe
  • Device Diagnostic can check the machine daily, weekly or every 2 weeks

Trade-offs

  • No Auto ramp setting: Ramp Time is Off or 5 to 45 minutes only
  • Ramp start pressure and all pressures are clinician settings, so you cannot fine-tune them from the patient menu
  • Only ClimateLineAir 11 heated tubing and HumidAir 11 tubs fit; AirCurve 10 parts do not
  • The standard tub needs distilled water only
  • Will not run without the tub or side cover in place
  • Typical power draw of about 65 W is higher than the AirSense 11, which shortens battery run time

Comfort settings you can change

These are in the patient menu. Your prescribed pressure and therapy mode are set by your clinician and are not covered here.

Ramp Time

Starts at a lower pressure and rises to your prescribed pressures over the set time.

How to change it: Tap MY OPTIONS, tap Ramp Time, choose Off or 5 to 45 minutes, then tap OK. The ramp start pressure is a clinician setting; if the start feels too low, ask your provider.

Humidity Level

Sets how much moisture the humidifier adds: up for a dry nose or mouth, down for water in the mask.

How to change it: Tap MY OPTIONS, swipe to Humidity Level, tap Off or 1 to 8, then tap OK.

Climate Control

With ClimateLineAir 11 and the tub attached, Auto holds about 85% relative humidity and guards against rainout; Manual lets you set temperature and humidity yourself.

How to change it: Tap MY OPTIONS, tap Climate Control, choose Auto or Manual, then tap OK.

Tube Temperature

Auto is 80 F (27 C); you can choose 60 to 86 F (16 to 30 C) or Off. Raising it is the first fix for rainout.

How to change it: Tap MY OPTIONS, tap Tube Temp, tap the setting, then tap OK.

Pressure Relief (EPR)

Eases breathing out in CPAP mode. In VAuto mode the gap between inhale and exhale pressure is your prescribed pressure support instead.

How to change it: If your care professional has enabled it and your mode supports it, tap MY OPTIONS, tap Pressure Relief and choose from the options shown. If you do not see it, that is set on the clinician side.

Mask

Selects the type of mask you use with the machine.

How to change it: Tap MY OPTIONS, tap Mask, select your mask type, then tap OK.

Tube

Selects the type of tubing attached.

How to change it: Tap MY OPTIONS, tap Tube, select your tubing, then tap OK.

SmartStart and SmartStop

Starts therapy when you breathe into the mask and stops it a few seconds after you take it off.

How to change it: If enabled by your care professional, tap MY OPTIONS and tap SmartStart or SmartStop to switch it on or off. SmartStop may not work with a conduit mask.

Run Warmup

Heats the water before you start so the air is not cold or dry.

How to change it: Tap MORE on the Home screen, then tap Run Warmup.

Airplane Mode

Turns off wireless transmission for flights.

How to change it: Tap MORE, swipe to Airplane Mode and tap it to switch on or off.

Cleaning and care

  1. Every day: empty and wipe the tub

    Let the tub cool for at least 10 minutes, or until the cool-down mode has finished. Remove it, pour out leftover water, wipe it with a clean disposable cloth and let it air dry out of direct sunlight. Refill before bed with the right water for your tub.

  2. Every week: wash tub, tubing and outlet connector

    Unplug the machine. Wash the tubing in warm water with mild dishwashing liquid. Wash the tub and outlet connector the same way or in 1 part vinegar to 9 parts water at room temperature.

    Keep water below 131 F (55 C), rinse well, dry out of direct sunlight, and wipe the machine's outside with a dry cloth.

  3. Outlet connector

    With the tub out, press the outlet connector's clip firmly and pull it out through the socket at the back. Refit it with the seal pointing left and the clip pointing forward.

  4. Dishwasher

    The tub and outlet connector may be washed in a dishwasher. The tubing must not go in a dishwasher or washing machine.

  5. What not to use

    Do not use bleach, chlorine, aromatic-based solutions, moisturizing or antibacterial soaps or scented oils. ResMed warns smoke, ozone and other gases may damage the machine and are not covered by the warranty.

  6. Air filter

    Replace the filter every six months, or sooner if it has holes or is clogged. It is not washable. Open the filter cover, swap the filter and close the cover; never run without both fitted.

  7. Inspect

    Replace the tub if it leaks, cracks, clouds or pits, or its seal tears; remove white deposits with 1 part vinegar to 9 parts water. Replace tubing with holes, tears or cracks.

Replacement parts

Tubing
ClimateLineAir 11 heated tubing (39102) or SlimLine tubing (36810).
Filters
Air11 standard filter (ResMed 39300 single, 39301 2-pack, 39302 12-pack) or hypoallergenic (39304, 39305, 39306). Not washable.
Water chamber
HumidAir 11 Standard tub (39100, distilled water) or HumidAir 11 Cleanable tub (39101, drinking-quality water); Air11 side cover (39226) for use without humidification.
Replacement notes
ResMed gives a design life of 6 months for the standard tub and tubing and 2.5 years for the cleanable tub, and says to replace the filter every six months. Medicare's PAP policy (LCD L33718) lists these usual maximum quantities: non-heated tubing (A7037) or heated tubing (A4604) 1 per 3 months, disposable filters (A7038) 2 per month, non-disposable filters (A7039) 1 per 6 months, and the humidifier water chamber (A7046) 1 per 6 months. These are the most Medicare usually pays for, not a rule that parts last that long; private insurers set their own schedules, so check with your plan and your equipment supplier.

Troubleshooting

Air leaking around the mask

Refit the mask, then run Mask Fit from the MORE menu or watch the Mask Fit video in myAir. Higher bilevel pressures make a good fit more important.

Too much air when you first lie down

Ramp may be off. Turn on Ramp Time in MY OPTIONS.

Not enough air at the start

Ramp may still be running; wait for pressure to build or turn Ramp Time off. If the start feels too low every night, ask your clinician about the ramp start pressure.

Water in the mask or tubing

Lower Humidity Level and raise Tube Temperature if you use ClimateLineAir 11. Keep the machine lower than your head.

Dry nose or mouth

Raise Humidity Level or use Climate Control Auto. A dry mouth often means mouth leak; ResMed suggests a chin strap or full face mask.

Still blowing air after you stop

Normal cool-down to avoid condensation; it stops by itself after 30 minutes.

Not getting enough air, or oxygen flow seems disrupted

Check the tubing for blockages, then remove and reassemble the tub correctly. If you use supplemental oxygen and this keeps happening, contact your care provider.

SmartStart does not start therapy

Take a deep breath in and out through the mask before breathing normally, or press the Start Therapy/Standby button. Fix any big leak or loose tubing first.

Data not sent or prescription not applied

Keep it plugged in where there is cellular signal, not on the floor or in a drawer, and turn off Airplane Mode. If data transfer is not enabled, your care provider needs to set it up.

Bloating or swallowing air

Note when it happens and tell your clinician; it is a listed side effect of PAP therapy, and pressure changes are their decision. Do not adjust pressures yourself.

Travel and power

ResMed states the AirCurve 11 meets FAA requirements for all phases of air travel, and it can go in the cabin as a medical device that does not count toward your carry-on limit; print a compliance letter from resmed.com. Empty the tub and pack it separately; on the plane, the empty tub or side cover must be in place to run, Airplane Mode should be on when crew ask, and the tub must never hold water in transit.

The 90W power supply takes 100 to 240 V at 50 to 60 Hz. For DC power ResMed lists the Air11 DC/DC converter (39231). Bilevel pressures plus heated humidification use more power than CPAP, so if you plan to use a battery, confirm with the battery maker that it supports the AirCurve 11 and expect better run time with humidity off. Cellular data may not transmit outside the country of purchase.

Your sleep data

MY SLEEP VIEW shows last night's hours, events per hour and mask seal, and myAir adds a nightly score and coaching; ResMed says myAir now supports bilevel users on AirCurve 11. Cellular data goes to your care provider after each session when the machine is plugged in and not in Airplane Mode.

Like the AirSense 11, the AirCurve 11 SD card slot sits behind a removable cover, and the guide says to remove the cover and insert a card to record your sleep data on it. Do not remove the card while its light is flashing. The free OSCAR program reads ResMed SD card data and its release notes mention ResMed 11 series machines. Bilevel data is more complex to read than CPAP data, so share anything that worries you with your clinician rather than changing settings.

Supplies that can help

Distilled water for CPAP humidifiers

Distilled water is what most CPAP humidifier chambers are meant to be filled with, because it leaves little or no mineral scale behind.

Replacement CPAP air filters

The air inlet filter keeps dust and lint out of the machine and your airway; it is a cheap part that needs regular replacement.

Heated CPAP tubing

Tubing with a built-in heating wire that keeps humidified air warm all the way to the mask, which is the main fix for rainout (water droplets …

Standard (unheated) CPAP tubing

The plain hose between the machine and the mask; it wears out and should be replaced on a regular schedule.

Humidifier water chamber (tub)

The removable water tank of a CPAP humidifier; it wears and needs replacing, especially if it cracks, clouds or leaks.

CPAP chinstrap

A soft strap that supports the jaw to help keep the mouth closed, used mainly with nasal and nasal pillow masks to reduce mouth leaks.

CPAP tubing cover

A fabric sleeve that slips over CPAP tubing to insulate it, reduce condensation and make the hose feel softer against your skin.

CPAP DC power cord or converter

A cord or converter that runs a CPAP from a 12 V or 24 V DC source, such as a car outlet, RV or deep cycle …

CPAP battery

A battery pack that runs a CPAP machine without mains power, for camping, power outages or long flights.

Mild unscented soap for CPAP cleaning

Plain, mild dish soap is the cleaning agent most CPAP makers and the FDA point to for masks, tubing and humidifier chambers.

Retailer links are affiliate links. We may earn a commission at no cost to you.

Frequently asked questions

What is the difference between AirCurve 11 VAuto and AirSense 11 AutoSet?

The AirSense 11 AutoSet delivers one auto-adjusting pressure up to 20 cm H2O, with EPR for a slight drop when you exhale. The AirCurve 11 VAuto is a bilevel machine with separate inhale and exhale pressures, pressure support up to 10 cm H2O and a ceiling of 25 cm H2O. Which one you need is your doctor's call.

Does the AirCurve 11 VAuto have AutoRamp?

No. ResMed's AirCurve 11 guide lists Ramp Time as Off or 5 to 45 minutes, with no Auto option. The ramp start pressure is set by your clinician.

Can I use AirCurve 10 tubing or tubs on the AirCurve 11?

No. The AirCurve 11 uses HumidAir 11 tubs and, for heated tubing, ClimateLineAir 11. SlimLine tubing (36810) is listed for the Air11 platform.

What is the AirCurve 11 S?

It is the fixed bilevel model in the same AirCurve 11 guide: it delivers set IPAP and EPAP pressures (IPAP 4 to 25, EPAP 3 to 25 cm H2O) instead of adjusting automatically. It shares the same body, humidifier and menus.

Why does my AirCurve 11 keep blowing air after I stop?

It is cooling down to keep condensation out of the tubing. ResMed says it stops on its own after 30 minutes.

Sources

Last reviewed October 11, 2026. How we research · Report an error