Philips Respironics · Nasal mask

Philips Respironics DreamWisp

A minimal contact nasal mask that combines the Wisp over the nose cushion with DreamWear's top of head tube. Air flows through the frame so the hose stays out of your face. Philips states that the headgear clips and mask frame contain magnets, and DreamWisp was part of the 2022 magnet correction.

This mask uses magnets

Philips states that the DreamWisp "headgear clips and mask frame contain magnets". Philips Respironics issued an Urgent Medical Device Correction dated August 29, 2022 for this mask, and the FDA published a safety communication about it on September 6, 2022 ("Certain Philips Respironics Masks for BiPAP, CPAP Machines Recalled Due to Safety Issue with Magnets That May Affect Certain Medical Devices"). The updated Philips contraindication reads: "Use of the mask is contraindicated for patients and their household members, caregivers, and bed partners that may be in close vicinity to patients using the masks, that have implanted devices that may be affected by magnets, including but not limited to": pacemakers; implantable cardioverter defibrillators (ICD); neurostimulators; magnetic metallic implants, electrodes or valves placed in the upper limbs, torso, or higher (neck and head); CSF (cerebral spinal fluid) shunts such as a VP shunt; aneurysm clips; embolic coils; intracranial aneurysm intravascular flow disruption devices; metallic cranial plates, screws, burr hole covers and bone substitute devices; metallic splinters in the eye; ocular implants (such as glaucoma or retinal implants); certain contact lenses with metal; implants to restore hearing or balance that have an implanted magnet (such as cochlear implants, implanted bone conduction hearing devices and auditory brainstem implants); magnetic denture attachments; metallic gastrointestinal clips; metallic stents (aneurysm, coronary, tracheobronchial, biliary); implantable ports and pumps (such as insulin pumps); hypoglossal nerve stimulators; devices labeled as MR (magnetic resonance) unsafe; and magnetic metallic implants not labeled for MR or not evaluated for safety in a magnetic field. The updated Philips warning reads: "Magnets with a magnetic field strength of 400 mT are used in the mask. With the exception of the devices identified in the contraindication, ensure the mask is kept at least 6 inches (approx. 15.24 cm) away from any other medical implants or medical devices that can be impacted by the magnetic fields." Philips reported 14 cases of possible interference with implanted devices, including pacemaker failure and defibrillator resets. If any of this applies to you or to anyone who sleeps near you, stop using the mask, call your doctor about a non-magnetic mask, and contact Philips Customer Care at 1-800-345-6443.

SizesCushion: Petite (P), Small (S), Medium (M), Large (L), Extra Large (XL), Frame connector: Small, Medium, Headgear: standard, large, and reduced size (RS), Fit pack with medium connector and Small, Medium, Large cushions
Works withAll standard 22 mm CPAP, APAP and BiPAP machines from any brand (ResMed, Philips, Fisher & Paykel, React Health and others). Select the matching mask type in your machine's comfort menu if it has one.

Usually suits

  • Wisp users who want the hose on top of the head
  • Side and stomach sleepers who breathe through the nose
  • People with no implants in the household

Design

  • Over the nose minimal contact cushion with an auto seal design
  • Top of head tubing through the frame
  • Magnetic headgear clips and magnetic frame
  • Open field of view
  • Operating pressure 4 to 30 cm H2O per Philips

Strengths and trade-offs

  • Hose on top of the head suits side and stomach sleepers
  • Small nasal cushion with a wide field of view
  • Magnetic clips make the headgear quick to attach and detach
  • Wide pressure range (4 to 30 cm H2O)
  • Magnets in both the clips and frame make it contraindicated for some implants in you or a bed partner
  • Cushion covers the nose tip, so it can touch the lower bridge area
  • Nasal only, so mouth breathing leaks

Finding your size

Like the Wisp, the cushion should fit the width of your nose without blocking your nostrils. The fit pack includes Small, Medium and Large cushions so you can compare them on your face, and Petite and Extra Large are sold separately. The frame comes with a small or medium connector and three headgear sizes, so ask your fitter to check those too if the mask does not sit right.

How to fit it

  1. Choose the cushion

    Pick the cushion that fits the width of your nose without blocking the nostrils. Start with the medium from the fit pack.

  2. Put it on

    Place the cushion over your nose, then pull the headgear over the top of your head.

  3. Clip the headgear

    Let the magnetic clips attach to the frame. Keep the mask at least 6 inches from anyone with an implanted medical device.

  4. Adjust and check

    Adjust the straps evenly without overtightening. Connect the hose at the top of your head, start the machine, and adjust lying down in your sleep position.

Common problems and fixes

Leak into the eyes

The cushion may be too large or sitting too high. Lower it toward the tip of the nose and recheck the size.

Mask shifts when you turn over

Check the headgear and connector size with your supplier. A CPAP pillow can reduce pressure on the frame.

Mouth leaks and dry mouth

Try a chinstrap. If mouth breathing continues, ask your prescriber about a full face option.

Clips detach during the night

Clean the magnet contact surfaces and make sure the straps are not twisted. Replace headgear that has stretched out.

Cleaning and care

Philips' instructions are to hand wash with a mild detergent, rinse thoroughly, and hang or air dry. Inspect the mask daily for wear and replace parts at the first sign of wear. Do not use bleach, alcohol, or ozone or UV cleaners.

How often to replace parts

Mask
1 every 3 months (Medicare A7034, nasal interface)
Cushion
Up to 2 per month (Medicare A7032, nasal cushion)
Headgear
1 every 6 months (Medicare A7035)

Get a free reminder when each part is due →

Supplies that can help

CPAP chinstrap

A soft strap that supports the jaw to help keep the mouth closed, used mainly with nasal and nasal pillow masks to reduce mouth leaks.

CPAP pillow

A pillow with cutouts or a contoured edge so a mask and hose are not pushed out of place when you sleep on your side.

Saline nasal spray

Salt water spray that moistens and clears the nose, a simple aid for the congestion and dryness some people get on CPAP.

Mild unscented soap for CPAP cleaning

Plain, mild dish soap is the cleaning agent most CPAP makers and the FDA point to for masks, tubing and humidifier chambers.

Retailer links are affiliate links. We may earn a commission at no cost to you.

Frequently asked questions

Does DreamWisp have magnets?

Yes. Philips says both the headgear clips and the mask frame contain magnets. DreamWisp was included in Philips' 2022 magnet correction, so it should not be used if you or anyone who sleeps near you has a listed implant.

What is the difference between DreamWisp and DreamWear Nasal?

DreamWisp uses a Wisp style cushion that covers the tip of the nose, while DreamWear Nasal cradles the underside of the nose. Both route air to the top of the head, but only DreamWisp has magnets.

What sizes does DreamWisp come in?

Cushions come in Petite, Small, Medium, Large and Extra Large, the frame connector comes in small and medium, and the headgear in standard, large and reduced size. The fit pack has Small, Medium and Large cushions.

Sources

Last reviewed October 11, 2026. How we research · Report an error