CPAP power draw (manufacturer figures)

These figures come from manufacturer documents and help you size a battery or check an inverter. Humidification and heated tubing are by far the biggest power users.

DeviceConfigurationPublished figureApprox. wattsBattery for 8 hours (12 V)
ResMed AirSense 11Typical, per user guide (24 V, 65 W power supply)56.1 W typical56Not published
ResMed AirSense 11Peak, per user guide73.2 W peak73Not published
ResMed AirMiniTypical, per fact sheet (24 V, 0.83 A output)6.3 W typical6Not published
ResMed AirMiniPeak, per fact sheet27 W peak (1 W standby)27Not published
ResMed AirSense 1010 cm H2O, SlimLine tubing, no humidifier0.93 A at 12 V DC1112 Ah
ResMed AirSense 1010 cm H2O, SlimLine tubing, humidifier at 42.30 A at 12 V DC2828 Ah
ResMed AirSense 1010 cm H2O, SlimLine tubing, humidifier at 83.66 A at 12 V DC4444 Ah
ResMed AirSense 1010 cm H2O, ClimateLineAir, humidifier on Auto3.77 A at 12 V DC4546 Ah
ResMed AirSense 1010 cm H2O, ClimateLineAir at 30 C, humidifier at 84.92 A at 12 V DC5960 Ah

Notes

  • Approximate watts are our arithmetic: for 12 V DC rows, 12 V times the published current; for other rows, the published wattage rounded.
  • AirSense 10 figures and battery sizes come from ResMed's battery guide (document 198103/4, dated August 2015), which uses the Air10 DC/DC converter; its battery sizes include a 50% safety margin.
  • Higher pressure, higher humidity, heated tubing and mask leaks all increase draw. Turning off the humidifier and heated tube is the most effective way to extend battery life.
  • We list only figures the manufacturer has published. Other machines are omitted where we could not find a manufacturer figure.
  • Check your own model's user guide; figures can differ by region and revision.